The Latest Information on Piezo Nanopositioning Systems is Available Here
P-722, P-723 Microscope Objective NanoPositioners & Scanners
Ordering Information:
P-722.00 PIFOC Objective Positioner & Scanner, 200 µm, Thread W0.8 x1/36"
P-722.07 PIFOC Objective Positioner & Scanner, 200 µm, Specify Thread (P-722.01 - .08)
P-722.10 PIFOC Objective Positioner & Scanner, 200 µm, LVDT, W0.8x1/36"
P-722.17 PIFOC Objective Positioner & Scanner, 200 µm, LVDT, Specify Thread (P-722.01 - .08)
P-723.00 PIFOC Objective Positioner & Scanner, 350 µm, Thread W0.8x1/36"
P-723.07 PIFOC Objective Positioner & Scanner, 350 µm, Specify Thread (P-722.01 - .08)
P-723.10 PIFOC Objective Positioner & Scanner, 350 µm, LVDT, W0.8x1/36"
P-723.17 PIFOC Objective Positioner & Scanner, 350 µm, LVDT, Specify Thread (P-722.01 - .08)
See "PZT Control Electronics" section for our comprehensive line of low noise, modular and OEM control electronics for computer and manual control.
TECHNICAL DATA: Models
P-722.00
P-722.10
P-723.00
P-723.10
Units
Max. objective diameter
39
39
39
39
mm
Open loop travel @ 0 to 100 V
200
200
350
350
µm ± 20 %
A2
Closed loop travel ³
-
200
-
350
µm
A5
Integrated feedback sensor
-
LVDT
-
LVDT
B
Closed / open loop ** resolution£
- / 2
10 / 2
- / 3
10 / 3
nm
C1
Closed loop linearity (typ.)
-
0.1
-
0.1
%
Full range repeatability (typ.)
±40
±60
nm
C3
Stiffness
0.38
0.38
0.32
0.32
N/µm ± 20 %
D1
Push/pull force capacity
(in operating direction)20 / 20
20 / 20
20 / 20
20 / 20
N
D3
Tilt (Q X, Q Y) (typ.)
3
3
5
5
µrad
E1
Electrical capacitance
7.8
7.8
11.1
11.1
µF ± 20 %
F1
* Dynamic operating current
coefficient (DOCC)4.8
4.8
3.9
3.9
µA/(Hz x µm)
F2
Unloaded resonant frequency
300
300
150
150
Hz ± 20 %
G2
Resonant frequency @ 150 g load
180
180
100
100
Hz ± 20 %
G3
Operating temperature range
- 20 to 80
- 20 to 80
- 20 to 80
- 20 to 80
° C
H2
Voltage connection
VL
VL
VL
VL
J1
Sensor connection
-
L
-
L
J2
Weight (with cables)
400
410
410
420
g ± 5 %
Body material
Al
Al
Al
Al
L
Recommended driver/controller
G, C, (A)
H, E
G, C, (A)
H, E
M
* Dynamic Operating Current Coefficient in µA per Hertz and µm.
Example: Sinusoidal scan of 30 µm at 10 Hz requires approximately 1.4 mA drive current (P-722).** Noise equivalent motion with E-503 amplifier.