Basic Designs of Piezoelectric Positioning Elements

Laminar Design (Contraction Type Actuator)
The active material in the laminar actuators consists of thin ceramic strips. The displacement of these actuators is perpendicular to the direction of polarization and electric field. When the volt-age is increased the strip contracts. The piezo strain coefficient d31 (negative!) describes the relative change in length. It is on the order of 50% of d33.

The maximum travel is a function of the length of the strips, while the number of strips arranged in parallel determines the stiffness and the stability of the element.

Displacement of a Piezo contraction actuator can be estimated by the following equation:

DL » d31*L * U/d       (4-25)

where
d31 = strain coefficient (deflection normal to polarization direction) [m/V]

L = length of the Piezo ceramics [m]

U = operating voltage [V]

d = thickness of one ceramic layer [m]

Example:
Laminar piezos are used in the P-280. P-282 Flexure Positioners click here for "Piezo Flexure NanoPositioners" section.


Tube Design


Examples:
PT-120 to PT-140 Piezo Tubes click here for "Piezo Actuators" section.

Monolithic ceramic tubes are yet another form of piezo actuator. Tubes are silvered inside and out and operate on the transversal piezo effect. When an electric voltage is applied between the outer and inner diameter, the tube contracts axially and radially. Axial contraction can be estimated by the following equation:

DL » d31*L * U/d       (4-26a)

where
d31 = strain coefficient (deflection normal to polarization direction) [m/V]

L = length of the Piezo ceramic tube [m]

U = operating voltage [V]

d = wall thickness [m]

Dd » d33 * U       (4-26b)

where
Dd = change in wall thickness [m]

d33 = strain coefficient (field and deflection in polarization direction) [m/V]

U = operating voltage [V]

The radial contraction is a superposition of several effects and cannot be ex-pressed by a simple equation.

When the outside electrode of a tube is separated into four 90° segments, differential drive voltage of opposing electrodes will lead to bending of one end (if the other end is clamped). Scanner tubes that flex in X and Y are widely used in scanning probe microscopes.

The scan range of a scanner tube is defined by
Dx » (2Ö2 * d31 * U * L2) / (p * ID * d)       (4-27)

Where
Dx = scan range in X and Y (for symmetrical electrodes) [m]

d31 = strain coefficient (deflection normal to polarization direction) [m/V]

U = operating voltage [V]

L = length [m]

ID = inner diameter [m]

d = wall thickness [m]

Tube actuators are not designed to withstand large forces. Application examples are scanning microscopy, ink jet printers etc.


Scanner Tube